Printer Friendly Version
Printer Friendly Version
Printer Friendly Version
A.2.02.10
Cardiac resynchronization therapy (CRT), which consists of synchronized pacing of the left and right ventricles, is intended to treat individuals with heart failure and dyssynchronous ventricular contractions. Treatment involves placement of a device that paces both ventricles and coordinates ventricular pacing to maximize cardiac pumping function and left ventricular ejection fraction.
Heart Failure
An estimated 6.7 million adults in the United States 20 years of age and older had heart failure between 2017 to 2020. The prevalence continues to increase over time with the aging of the population. Prevalence of disease is higher in women than men 80 years of age and older. Overall prevalence is especially high in Black individuals. A 2008 study demonstrated that Black individuals had the highest risk of developing heart failure, followed by Hispanic, White, and Chinese individuals in the United States. Higher risk reflected differential prevalence of hypertension, diabetes, and lower socioeconomic status. Black individuals also had the highest proportion of incident heart failure not preceded by myocardial infarction (75%). Additionally, Black individuals have a greater 5-year case fatality rate associated with heart failure compared to White individuals. It is estimated that 20% to 30% of patients with heart failure have intraventricular conduction disorders resulting in a contraction pattern that is not coordinated and a wide QRS interval on the electrocardiogram. This abnormality appears to be associated with increased morbidity and mortality.
TreatmentBiventricular pacemakers using three leads (one in the right atrium, one endocardial in the right ventricle, and one epicardial for the left ventricle), also known as cardiac resynchronization therapy (CRT), have been investigated as a technique to coordinate the contraction of the ventricles, thus improving patients' hemodynamic status. Originally developed CRT devices typically used two ventricular leads for biventricular pacing. Devices and implantation techniques have been developed to allow for multisite pacing, with the goal of improving CRT response. This may be accomplished in one of two ways: through the use of multiple leads within the coronary sinus (triventricular pacing) or through the use of multipolar left ventricular pacing leads, which can deliver pacing stimuli at multiple sites. Wireless left ventricular endocardial pacing is also being evaluated for patients who are not candidates for or do not respond to standard epicardial pacing leads.
There are numerous CRT devices, combined implantable cardioverter-defibrillator (ICD) plus CRT devices (CRT-D), and combined CRT plus fluid monitoring devices. Some devices are discussed here. For example, in 2001, the InSync® Biventricular Pacing System (Medtronic), a stand-alone biventricular pacemaker, was approved by the U.S. Food and Drug Administration (FDA) through the premarket approval process for the treatment of patients with New York Heart Association (NYHA) Class III or IV heart failure, on a stable pharmacologic regimen, who also have a QRS duration of 130 ms or longer and a left ventricular ejection fraction of 35% or less. Devices by Guidant (CONTAK-CD® CRT-D System) and Medtronic (InSync® ICD Model 7272) have been approved by the FDA through the premarket approval process for combined cardiac resynchronization therapy defibrillators for patients at high risk of sudden cardiac death due to ventricular arrhythmias and who have NYHA Class III or IV heart failure with a left ventricular ejection fraction of 35% or less, QRS interval 130 ms or longer (≥120 ms for the Guidant device), and remain symptomatic despite a stable, optimal heart failure drug therapy. In 2006, Biotronik Inc. received premarket approval from the FDA for its combined CRT-D device with ventricular pacing leads (Tupos LV/ATx CRT-D/Kronos LV-T CRT-D systems); in 2013, the company received the FDA approval for updated CRT-D devices (Ilesto/Iforia series). On the basis of the Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy (MADIT-CRT) study, indications for 3 Guidant CRT-D (Cognis®, Livian®, and Contak Renewal; Boston Scientific) devices were expanded to include patients with heart failure who receive stable optimal pharmacologic therapy for heart failure and who meet any one of the following classifications:
Moderate-to-severe heart failure (NYHA class III or IV) with an ejection fraction less than 35% and QRS interval greater than 120 ms.
Left bundle branch block with a QRS interval greater than or equal to 130 ms, ejection fraction less than 30%, and mild (NYHA class II) ischemic or nonischemic heart failure or asymptomatic (NYHA class I) ischemic heart failure.
In April 2014, the FDA further expanded indications for multiple Medtronic CRT devices to include patients with NYHA class I, II, or III heart failure, who have an LVEF of 50% or less on stable, optimal heart failure medical therapy, if indicated, and have atrioventricular block that is expected to require a high percentage of ventricular pacing that cannot be managed with algorithms to minimize right ventricular pacing. The expanded indication was based on data from the Biventricular versus Right Ventricular Pacing in Heart Failure Patients with Atrioventricular Block (BLOCK-HF) study, a Medtronic-sponsored randomized controlled trial that evaluated the use of CRT in patients with NYHA class I, II, or III heart failure, LVEF of 50% or less, and atrioventricular block.
Several CRT devices have incorporated a fourth lead, providing quadripolar pacing. The Medtronic Viva™ Quad XT and the Viva Quad S have a fourth lead, and the Medtronic Attain Performa® has a left ventricular lead, which received clearance for marketing from the FDA in August 2014. The Dynagen™ X4 and Inogen™ X4 devices (Boston Scientific) also incorporate a fourth lead. Other CRT devices with quadripolar leads have been approved for use outside of the United States (eg, St. Jude Quartet™ left ventricular lead).
Multiple devices manufactured by Medtronic combine a CRT with the OptiVol™ monitoring system. For example, in 2005, the InSync Sentry® system was approved by the FDA through the supplemental premarket approval process. This combined biventricular pacemaker plus ICD is also equipped to monitor intrathoracic fluid levels using bioimpedance technology, referred to as OptiVol™ Fluid Status Monitoring. Bioimpedance measures, defined as the electrical resistance of tissue to flow of current, are performed many times a day using a vector from the right ventricular coil on the lead in the right side of the heart to the implanted pacemaker devices; changes in bioimpedance reflect intrathoracic fluid status and are evaluated using a computer algorithm. For example, changes in a patient's daily average of intrathoracic bioimpedance can be monitored; differences in the daily average are compared with a baseline and reported as the OptiVol™ Fluid Index. It has been proposed that these data may be used as an early warning system of cardiac decompensation, or may provide feedback that enables a physician to tailor medical therapy. The Cardiac Hemodynamic Monitoring for the Management of Heart Failure in the Outpatient Setting medical policy addresses the use of external bioimpedance devices as stand-alone devices to assess cardiac output noninvasively.
The WiSE-CRT (EBR Systems) provides CRT with a small wireless electrode that is implanted within the left ventricle and controlled by ultrasound. It has European CE approval and is being studied in a multicenter pivotal trial. The final submission of its FDA premarket approval application was submitted in late 2024.
Biventricular pacemakers with or without an accompanying implantable cardiac defibrillator (i.e., a combined biventricular pacemaker plus implantable cardiac defibrillator) may be considered medically necessary as a treatment of heart failure in individuals who meet all of the following criteria:
NYHA Class III or IV
Left ventricular ejection fraction <35%
Sinus rhythm
Individuals treated with guideline-directed medical therapy (see Policy Guidelines section) AND
Either left bundle branch block OR QRS interval ≥150 ms.
NYHA Class II
Left ventricular ejection fraction ≤30%
Sinus rhythm
Individuals treated with a guideline-directed medical therapy (see Policy Guidelines section) AND
Either left bundle branch block OR QRS interval ≥150 ms.
For individuals who do not meet the criteria outlined above, but have an indication for a ventricular pacemaker, biventricular pacemakers with or without an accompanying implantable cardiac defibrillator (ie, a combined biventricular pacemaker plus implantable cardiac defibrillator) may be considered medically necessary as an alternative to a right ventricular pacemaker in individuals who meet all of the following criteria:
NYHA class I, II, III, or IV heart failure;
Left ventricular ejection fraction ≤50%;
The presence of atrioventricular block with requirement for a high percentage of ventricular pacing (see Policy Guidelines section); and
Individuals treated with guideline-directed medical therapy (see Policy Guidelines section).
The New York Heart Association (NYHA) functional class is a widely used rating system that quantifies the degree of symptom severity. Each class indicates a value that is reflective of the cardiac functional status. The different functional stages are as follows:
New York Heart Association (NYHA) Functional Class | |
Class | Description |
I | No limitations; Ordinary physical activity does not cause symptoms |
II | Slight limitations; Comfortable at rest. Ordinary physical activity causes symptoms |
III | Marked limitations; Less than ordinary activity causes symptoms |
IV | Class IV Inability to carry out any physical activity; symptoms present at rest |
Andreoli et al., 1993 |
Biventricular pacemakers, with or without an accompanying implantable cardiac defibrillator (i.e., a combined biventricular pacemaker plus implantable cardiac defibrillator), are considered investigational as a treatment for individuals with NYHA class I heart failure who do not meet the above criteria.
Biventricular pacemakers, with or without an accompanying implantable cardiac defibrillator (i.e., a combined biventricular pacemaker plus implantable cardiac defibrillator), are considered investigational as a treatment for heart failure in individuals with atrial fibrillation who do not meet the above criteria.
Triple-site (triventricular) cardiac resynchronization therapy, using an additional pacing lead, is considered investigational.
An intrathoracic fluid monitoring sensor is considered investigational as a component of a biventricular pacemaker.
Cardiac resynchronization therapy with wireless left ventricular endocardial pacing is considered investigational.
None
The coverage guidelines outlined in the Medical Policy Manual should not be used in lieu of the Member's specific benefit plan language.
Definitions
Atrioventricular block with a requirement for a high percentage of ventricular pacing is considered to be present when there is either:
Third-degree atrioventricular block; or
Second-degree atrioventricular block or a PR interval of 300 ms or more when paced at 100 beats per minute.
Guideline-directed medical therapy for heart failure is outlined in the 2022 American Heart Association, American College of Cardiology, and Heart Failure Society of America guidelines for the management of heart failure.
Medically Necessary is defined as those services, treatments, procedures, equipment, drugs, devices, items or supplies furnished by a covered Provider that are required to identify or treat a Member's illness, injury or Mental Health Disorders, and which Company determines are covered under this Benefit Plan based on the criteria as follows in A through D:
A. consistent with the symptoms or diagnosis and treatment of the Member's condition, illness, or injury; and
B. appropriate with regard to standards of good medical practice; and
C. not solely for the convenience of the Member, his or her Provider; and
D. the most appropriate supply or level of care which can safely be provided to Member. When applied to the care of an Inpatient, it further means that services for the Member's medical symptoms or conditions require that the services cannot be safely provided to the Member as an Outpatient.
For the definition of medical necessity, “standards of good medical practice” means standards that are based on credible scientific evidence published in peer-reviewed medical literature generally recognized by the relevant medical community, and physician specialty society recommendations, and the views of medical practitioners practicing in relevant clinical areas and any other relevant factors. BCBSMS makes no payment for services, treatments, procedures, equipment, drugs, devices, items or supplies which are not documented to be Medically Necessary. The fact that a Physician or other Provider has prescribed, ordered, recommended, or approved a service or supply does not in itself, make it Medically Necessary.
Investigative is defined as the use of any treatment procedure, facility, equipment, drug, device, or supply not yet recognized as a generally accepted standard of good medical practice for the treatment of the condition being treated and; therefore, is not considered medically necessary. For the definition of Investigative, “generally accepted standards of medical practice” means standards that are based on credible scientific evidence published in peer-reviewed medical literature generally recognized by the relevant medical community, and physician specialty society recommendations, and the views of medical practitioners practicing in relevant clinical areas and any other relevant factors. In order for equipment, devices, drugs or supplies [i.e, technologies], to be considered not investigative, the technology must have final approval from the appropriate governmental bodies, and scientific evidence must permit conclusions concerning the effect of the technology on health outcomes, and the technology must improve the net health outcome, and the technology must be as beneficial as any established alternative and the improvement must be attainable outside the testing/investigational setting.
2/2000: Approved by Medical Policy Advisory Committee (MPAC).
2/8/2002: Investigational definition added.
4/18/2002: Type of Service and Place of Service deleted.
8/2002: Reviewed by MPAC; investigational status changed to medically necessary.
12/16/2002: New York Heart Association (NYHA) functional class added, Code Reference section updated, CPT code 33224, 33225, 33226, 33999 added, ICD-9 diagnosis code 402.01, 402.11, 402.91, 404.01, 404.03, 404.11, 404.13, 404.91, 404.93, 428.0, 428.9 added.
3/4/2003: CPT code 33215 added.
8/29/2003: "Description" section updated to be consistent with BCBSA.
8/9/2005: Code Reference section updated, CPT code 33215, 33999 deleted, ICD-9 procedure code 00.52 added.
9/13/2006: Coding revised. ICD9 2006 revisions added to policy.
12/13/2006: Policy reviewed, intrathoracic fluid monitoring noted to be investigational.
2/19/2008: Policy reviewed, no changes.
01/14/2010: Description section updated, Policy Statement was updated to include that biventricular pacemakers with or without an accompanying implantable cardiac defibrillator (i.e., a combined biventricular pacemaker/ICD) may be considered medically necessary based on indications. Also corrected typographic error, changed QRS duration from >20 msec to >120 msec. Added that biventricular pacemakers with or without an accompanying implantable cardiac defibrillator are considered investigational as a treatment of NYHA class I or II heart failure. Coding Section updated to add the following verbiage: "*Some covered procedure codes may have multiple descriptions. Coverage will only be made for covered codes when used for services outlined within the policy statement section." Removed ICD-9 Procedure code 00.52 and added "unspecified" to ICD-9 diagnosis codes 428.0 and 428.9 description.
04/28/2010: Policy description updated to add link to related medical policy. Clarified the policy statement regarding biventricular pacemakers, with or without an accompanying implantable cardiac defibrillator; intent unchanged. Deleted outdated references from the Sources section.
08/03/2011: Policy statement revised to add specific coverage criteria for NYHA class II heart failure. Deleted "congestive" from the policy title and statement.
07/12/2012: Added Cardiac Resynchronization Therapy to the policy title. Deleted "(or angiotensin receptor blocker)" from the fourth bullet after beta blocker in both of the NYHA criteria lists. Added "sinus rhythm" to the class III or IV statement for clarity purposes.
09/25/2012: Added the following investigational policy statement: Biventricular pacemakers, with or without an accompanying implantable cardiac defibrillator (i.e., a combined biventricular pacemaker/ICD), are considered investigational as a treatment for heart failure in patients with atrial fibrillation.
05/06/2013: Added the following investigational policy statement: Triple-site (triventricular) CRT, using an additional pacing lead, is considered investigational.
07/16/2014: Policy reviewed; description updated regarding devices. Revised policy statement to state that the FDA-labeled indications for QRS duration vary by device. For some devices, FDA approval is based on QRS duration of >130 (eg, InSync® device), while for others, it is based on QRS duration >120 ms (eg, CONTAK CD® CRT-D System). These differences in QRS duration arise from differences in the eligibility criteria in the trials on which the FDA approval is based.
12/31/2014: Code Reference section updated to revise the description of the following CPT codes: 33224 and 33225.
08/21/2015: Code Reference section updated for ICD-10.
06/08/2016: Policy number A.2.02.10 added. Policy Guidelines updated to add medically necessary and investigative definitions.
08/24/2016: Policy description updated regarding indications for CRT devices. Medically necessary criteria for NYHA Class II and III/IV updated with the following changes: 1) Third bullet point changed to "patients with guideline-directed medical therapy." 2) Updated criteria to include presence of left bundle branch block OR QRS >150 ms. Added policy statement that for patients who have an indication for a ventricular pacemaker, biventricular pacemakers with or without an accompanying implantable cardiac defibrillator may be considered medically necessary as an alternative to a right ventricular pacemaker with criteria. Policy guidelines updated to add policy definitions.
06/01/2017: Policy reviewed; no changes.
09/29/2017: Code Reference section updated to add new ICD-10 diagnosis codes I50.810, I50.811, I50.812, I50.813, I50.814, I50.82, I50.83, I50.84, and I50.89. Effective 10/01/2017.
05/23/2018: Medical policy links updated in policy description.
08/02/2018: Policy description updated regarding wireless left ventricular endocardial pacing and devices. First medically necessary statement updated to change "duration" to "interval." Added statement that cardiac resynchronization therapy with wireless left ventricular endocardial pacing is considered investigational.
06/04/2019: Policy description updated. Policy statements unchanged.
06/12/2020: Policy reviewed; no changes.
07/26/2021: Policy description updated. Policy statements unchanged. Policy Guidelines updated to change "Nervous/Mental Conditions" to "Mental Health Disorders" and "Medically Necessary" to "medical necessity."
06/08/2022: Policy description updated regarding heart failure. Policy statements updated to change "patients" to "individuals." Policy intent unchanged. Policy Guidelines updated regarding guidelines for the management of heart failure.
06/20/2023: Policy description updated regarding adults with heart failure. Second policy statement revised to state: For individuals who do not meet the criteria outlined above, but have an indication for a ventricular pacemaker, biventricular pacemakers with or without an accompanying implantable cardiac defibrillator (ie, a combined biventricular pacemaker plus implantable cardiac defibrillator) may be considered medically necessary as an alternative to a right ventricular pacemaker in individuals who meet all of the following criteria. It previously stated: For individuals who do not meet the criteria outlined above, but have an indication for a ventricular pacemaker or biventricular pacemakers with or without an accompanying implantable cardiac defibrillator (ie, a combined biventricular pacemaker/implantable cardiac defibrillator) may be considered medically necessary as an alternative to a right ventricular pacemaker in individuals who meet all of the following criteria.
12/21/2023: Code Reference section updated to add new 2024 CPT codes 0861T, 0862T, and 0863T, effective 01/01/2024.
06/11/2024: Policy description updated. Policy statements unchanged.
08/07/2025: Policy description updated regarding devices. Policy statements unchanged.
Blue Cross Blue Shield Association policy # 2.02.10
This may not be a comprehensive list of procedure codes applicable to this policy.
Code Number | Description | ||
CPT-4 | |||
33224 | Insertion of pacing electrode, cardiac venous system, for left ventricular pacing, with attachment to previously placed pacemaker or implantable defibrillator pulse generator (including revision of pocket, removal, insertion, and/or replacement of existing generator) | ||
33225 | Insertion of pacing electrode, cardiac venous system, for left ventricular pacing, at time of insertion of implantable defibrillator or pacemaker pulse generator (eg, for upgrade to dual chamber system) (List separately in addition to code for primary procedure) Use 33225 in conjunction with 33206, 33207, 33208, 33212, 33213, 33214, 33216, 33217, 33222, 33233, 33234, 33235, 33240, 33249) | ||
33226 | Repositioning of previously implanted cardiac venous system (left ventricular) electrode (including removal, insertion and/or replacement of generator) | ||
HCPCS | |||
ICD-9 Procedure | ICD-10 Procedure | ||
ICD-9 Diagnosis | ICD-10 Diagnosis | ||
402.01 | Malignant hypertensive heart disease with congestive heart failure | I11.0 | Hypertensive heart disease with heart failure *NOTE: Use additional code to identify type of heart failure (I50.-) |
402.11 | Benign hypertensive heart disease with congestive heart failure | ||
402.91 | Unspecified hypertensive heart disease with congestive heart failure | ||
404.01 | Hypertensive heart and chronic kidney disease, malignant, with heart failure and with chronic kidney disease stage I through stage IV, or unspecified | I13.0 | Hypertensive heart and chronic kidney disease with heart failure and stage 1 through 4 chronic kidney disease, or unspecified chronic kidney disease *NOTE: Use additional code to identify type of heart failure (I50.-) Use additional code to identify stage of chronic kidney disease (N18.1-N18.4, N18.9) |
404.11 | Hypertensive heart and chronic kidney disease, benign, with heart failure and with chronic kidney disease stage I through stage IV, or unspecified | ||
404.91 | Hypertensive heart and chronic kidney disease, unspecified, with heart failure and with chronic kidney disease stage I through stage IV, or unspecified. | ||
404.03 | Hypertensive heart and chronic kidney disease, malignant, with heart failure and with chronic kidney disease stage V or end stage renal disease | I13.2 | Hypertensive heart and chronic kidney disease with heart failure and stage 5 chronic kidney disease, or end stage renal disease *NOTE: Use additional code to identify type of heart failure (I50.-) Use additional code to identify stage of chronic kidney disease (N18.5, N18.6) |
404.13 | Hypertensive heart and chronic kidney disease, benign, with heart failure and with chronic kidney disease stage V or end stage renal disease | ||
404.93 | Hypertensive heart and chronic kidney disease, unspecified, with heart failure and with chronic kidney disease stage V or end stage renal disease. | ||
428.0 | Congestive heart failure, unspecified | I50.20-I50.9 | Heart failure (code range) |
428.9 | Unspecified heart failure, unspecified | I50.9, I50.810, I50.811, I50.812, I50.813, I50.814, I50.82, I50.83, I50.84, I50.89 | Heart failure |
Investigational Codes
Code Number | Description |
CPT-4 | |
0861T | Removal of pulse generator for wireless cardiac stimulator for left ventricular pacing; both components (battery and transmitter) |
0862T | Relocation of pulse generator for wireless cardiac stimulator for left ventricular pacing, including device interrogation and programming; battery component only |
0863T | Relocation of pulse generator for wireless cardiac stimulator for left ventricular pacing, including device interrogation and programming; transmitter component only |
HCPCS | |
ICD-10 Procedure | |
ICD-10 Diagnosis |
CPT copyright American Medical Association. All rights reserved. CPT is a registered trademark of the American Medical Association.